MODEL
|
7730
|
AC WITHSTAND VOLTAGE
|
|
Range
|
Resolution
|
Accuracy
|
Output Voltage, ACV
|
0-5000
|
1
|
±(2% of setting+ 5V)
|
Output Frequency
|
50Hz/60Hz ±0.1%, User Selection
|
Output Waveform
|
Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5
|
Output Regulation
|
± (1% of output + 5V), From no load to full load
|
SETTINGS
|
Maximum Limit current, mA
|
0.000 - 9.999 (0=OFF)
|
0.001
|
± (2% of setting + 2 counts)
|
10.00-30.00
|
0.01
|
Minimum Limit current, mA
|
0.000 - 9.999 (0=OFF)
|
0.001
|
± (2% of setting + 2 counts)
|
10.00-30.00
|
0.01
|
Ramp-Up Timer, second
|
0.1 - 999.9
|
0.1
|
± (0.1% + 0.05 sec)
|
Ramp-Down Timer, second
|
0.0 - 999.9
|
Dwell Timer, second
|
0, 0.4 - 999.9
(0=continuous)
|
Arc Detection
|
1 - 9 ranges (9 is the most sensitivity)
|
DC WITHSTAND VOLTAGE
|
Output Voltage, DCV
|
0-5000
|
1
|
±(2% of setting+ 5V)
|
Output Ripple
|
<4 %(5KV/10mA at Resistive Load)
|
SETTINGS
|
Maximum Limit, current, µA
|
0 - 999.9
|
0.1
|
± (2% of setting + 2 counts)
|
1000 - 10000
|
1
|
Minimum Limit, current, µA
|
0.0 - 999.9
|
0.1
|
± (2% of setting + 2 counts)
|
|
1000 - 10000
|
1
|
Ramp-Up Timer, second
|
0.1 - 999.9
|
0.1
|
± (0.1% + 0.05 sec)
|
Ramp-Down Timer, second
|
0, 1.0 - 999.9 (0=OFF)
|
Dwell Timer, second
|
0, 0.4 - 999.9
(0=continuous)
|
Ramp-HI, current
|
>12 mA peak maximum, ON/OFF User Selection
|
Charge-LO, current
|
0.0 - 350.0 µA DC or Auto Set
|
Discharge Time
|
≦ 200 msec
|
Maximum Capacitive LoadDC Mode
|
1 µF < 1KV 0.08 µF < 4KV
|
0.75 µF < 2KV 0.04 µF < 5KV
|
0.5 µF < 3KV
|
Arc Detection
|
1 - 9 ranges (9 is the most sensitivity)
|
INSULATION RESISTANCE
|
Output Voltage, VDC
|
50 - 1000
|
1
|
± (2% of setting + 2 counts)
|
Short Circuit Current
|
Maximum >12mA peak
|
SETTINGS
|
Maximum and Minimun Limit
Resistance, MΩ
|
0.05 - 99.99(Max – Limit: 0 = OFF)
|
0.01
|
Same as Resistance MEASUREMENT Accuracy
|
100.0-999.9
|
0.1
|
1000-50000
|
1
|
± (0.1% + 0.05 sec)
|
Ramp-Up Timer, second
|
0.1 - 999.9
|
0.1
|
Ramp-Down Timer, second
|
0, 1.0 - 999.9 (0=OFF)
|
Delay Timer, second
|
0, 1.0 - 999.9
(0=continuous)
|
0.1
|
Charge-LO, current, µA
|
0.000 - 3.500 or Auto Set
|
MEASUREMENT
|
|
Range
|
Resolution
|
Accuracy
|
Voltage, KV (AC/DC)
|
0.00-5.00
|
0.01
|
± (1.5% of reading +2counts)
|
Voltage, Vdc (IR only)
|
0-1000
|
1
|
± (1.5% of reading + 2counts)
|
AC Current, mA
|
0.000-3.500
|
0.001
|
± (2% of reading + 2counts)
|
3.00-30.00
|
0.01
|
DC Current, µA
|
0.0 - 350.0
|
0.1
|
± (2% of reading + 2 counts)
|
DC Current, mA
|
0.300 - 3.500
|
0.001
|
3.00-9.99
|
0.01
|
Resistance, MΩ
|
50-499V
|
500-1000V
|
|
50 – 499V
0.05–999.9, ±(7% of reading+2 counts)
|
0.050-1.999
|
0.050-9.999
|
0.001
|
500 – 1000V
|
2.00-19.99
|
10.00-99.99
|
0.01
|
0.10–999.9, ±(2% of reading+2 counts)
|
20.0-199.9
|
100.0-999.9
|
0.1
|
1000–9999, ±(5% of reading+2 counts)
|
200-50000
|
1000-50000
|
1
|
10000–50000, ±(15%of reading+2 counts)
|
GENERAL
|
Input Voltage
|
115/230Vac±15% Auto Selection, 50/60Hz ± 5%, Fuse 4A Slow-Blo 250VAC
|
PLC Remote Control
|
Input: Test, Reset, Interlock, Recall Memory 1 through 10
|
Output: Pass, Fail, Test-in-Process, Reset Out
|
Remote Output Relays
|
125VAC@1AAC, 30VDC@ 0.5ADC
|
Memory
|
50 memories, 30 steps/memory
|
Graphic Display
|
240 x 64 dot resolution Monographic LCD /Contrast 9 Levels 1-9
|
Safety
|
Built-in Smart GFI circuit,GFI trip current 450µA max, HV shut down speed: <1mS
|
Interface
|
Standard RS232, Optional PrinterPort,GPIB (IEEE-488.2),Multi-function Interface Card,Ethernet Card.
|
Security
|
Programmable password lockout capability to avoid unauthorized access to test set-up program.
|
Alarm Volume Setting
|
Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume.
|
Calibration
|
Adjustments are made through the fornt panel. Built-in Verification program.
|
Automatic Calibration alert function to signal operator when calibration is due.
|
Environment
|
0-40℃, 20-80%RH
|
Dimensions/Net Weight
|
(W × H × D) 2U (430 × 89 × 400 mm)/13Kg.
|
Option (Opt.735)
|
DC Continuity Test Function (0.1A/ 0-2KΩ) at Rear Panel
|
STANDARD ACCESSORIES
|
Power Cord(10A)
|
×1
|
Fuses
|
×2 (Including a spare contained in the fuse holder)
|
Interlock Disable Key(1505)
|
×1
|
Hipot Test Lead ,1.5m(1101)
|
×1
|
Hipot Return Lead ,1.8m(1102)
|
×1
|
RS232 Link Cable 1.5m(1130)
|
×1
|
MATRIX SCANNER MODULE (Opt.729)
|
High Voltage Rating
|
5KVAC/5KVDC
|
Number of HV Channel
|
8
|
MATRIX SCANNER MODULE (Opt.730)
|
High Voltage Rating
|
5KVAC/5KVDC
|
Number of HV Channel
|
4
|
DC CONTINUITY TEST FUNCTION (Opt.735)
|
MEASUREMENT
|
|
Range
|
Resolution
|
Accuracy
|
Output Current
|
DC 0.1A ± 0.01A, Max.
|
Resistance Display
|
0.00 – 19.99
|
0.01
|
± (1 % of reading + 5 counts)
|
20.0 – 199.9
|
0.1
|
± (1 % of reading + 2 counts)
|
200 – 2000
|
1
|
± (1 % of reading + 2 counts)
|
SETTINGS
|
Max-Limit, Ω
|
0.00 – 19.99
|
0.01
|
± (1 % of setting + 5 counts)
|
20.0 – 199.9
|
0.1
|
± (1 % of setting + 2 counts)
|
200 – 2000
|
1
|
± (1 % of setting + 2 counts)
|
Min-Limit, Ω
|
0.00 – 19.99
|
0.01
|
± (1 % of setting + 5 counts)
|
20.0 – 199.9
|
0.1
|
± (1 % of setting + 2 counts)
|
200 – 2000
|
1
|
± (1 % of setting + 2 counts)
|
Dwell Timer, second
|
0.0, 0.3 – 999.9
(0 = continuous)
|
0.1
|
± (0.1% + 0.05 sec)
|