Bandwidth
|
DC to 100MHz (-3dB)
|
Attenuation Ratio
|
1:100/1000
|
Accuracy
|
+/- 2%
|
Rise Time
|
3.5ns
|
Input Impedance
|
4Mohm//7pF each side to ground
|
Input Voltage
|
-Category
|
CAT III
|
-Differential Range
|
+/- 140V(DC+AC Peak ) and 140Vrms @ 1/100
+/-1400V(DC+AC Peak) and 1000Vrms@1/1000
|
-Common Mode Range
|
+/-1400V(DC+AC Peak) and 1000Vrms @1/100 & 1/1000
|
-Absolute Max. Voltage
(Differential or Common Mode)
|
+/-1400V(DC+AC Peak) and 1000Vrms @1/100 & 1/1000
|
Output Voltage
|
- Swing
|
+/-1.4V (into 50kohm load)
|
-Offset (typical)
|
<+/-5mV
|
-Noise (typical)
|
0.9mVrms
|
-Source Impedance (typical)
|
50 ohm (for using 1Mohm input system oscilloscope)
|
CMRR (typical)
|
-80dB @60Hz, -50dB @1MHz
|
Ambient Operating Temperature
|
-10 to 40 degree centigrade
|
Ambient Storage Temperature
|
-30 to 70 degree centingrade
|
Ambient Operating Humidity
|
25 to 85% RH
|
Ambient Storage Humidity
|
25 to 85% RH
|
Power Requirements*
|
-Standard
|
4xAA cells
|
-Options
|
Power leads, Mains adaptor* (6VDC/200mA or regulated 9VDC/120mA), USB power cord
|
length of Input Leads
|
30cm
|
Length of BNC Cable
|
92.5cm
|
Weight
|
500gms
|
Dimension(LxWxH)
|
207mm x 83mm x 38mm
|